How-To Videos

Need assistance on how to operate and maintain our products? Here you will find instructional videos for all of your needs. Of course you can also always reach out to your local distributor with our distributor map below!
Baldus Touch Nitrous Oxide

Baldus Touch:
Full Sedation Procedure

Learn how to perform a complete nitrous oxide sedation procedure using the Baldus Touch system. This video covers device startup with self-test, patient ID entry, total flow calculation based on patient height, stepwise N2O concentration adjustment up to 70%, continuous pulse oximeter monitoring, O2 flash function usage, and the 5-minute 100% oxygen elimination phase after treatment.

Baldus Touch:
System Bleeding Procedure – Correct Shutdown After Sedation

Learn how to correctly bleed the Baldus Touch nitrous oxide sedation system after use. This video covers the full shutdown sequence — closing N2O and O2 cylinders, monitoring pressure drop on the display, recognizing the “Check N2O/O2 gas supply” warning messages, and powering down the device via standby or full shutdown.

Baldus Analog:
Full Sedation Procedure

Learn how to perform a complete nitrous oxide sedation procedure using the Baldus Analog system. This video covers total flow adjustment using the right rotary knob, nitrous oxide concentration setting up to 70% using the left rotary knob, correct and gentle knob handling to prevent damage, and use of the O2 flush function for immediate delivery of 50–60 L/min pure oxygen.

Baldus Analog:
System Bleeding Procedure – Correct Shutdown After Sedation

Learn how to correctly bleed the Baldus Analog nitrous oxide sedation system after use. This video shows the full shutdown sequence — setting maximum N2O concentration, closing the N2O and O2 cylinders in order, observing the ball indicator drops on the flowmeter, and confirming complete bleeding via the system warning signal.

Gas Cylinder Change

Learn how to safely change gas cylinders on the Baldus Analog nitrous oxide sedation system. Example with German gas cylinders. This step-by-step guide covers protective cap removal, correct pressure regulator assignment by thread size, proper cart placement, color-coded hose connections, and safe cylinder opening — essential for any dental practice using N2O sedation.

Pressure Regulators Explained – Maintenance & Safe Cylinder Change

Learn how to correctly identify, maintain, and replace the pressure regulators on the Baldus Analog nitrous oxide sedation system. This video covers the difference between the N2O and O2 regulators, how thread sizes prevent mix-ups, and the critical 10-minute resting rule for the N2O seal before changing cylinders. Essential knowledge for annual maintenance and safe gas cylinder handling in dental practices.

Scavenger System Assembly

Learn how to correctly assemble the Baldus Scavenger System for nitrous oxide sedation in dental practices. This video walks through every component — breathing bag, corrugated hose, Beck tee connector, vacuum control block, and suction hose — and shows the exact assembly sequence and correct port assignments.

Vacuum Control Block Setup & Suction Adjustment

Learn how to connect and correctly adjust the vacuum control block on the Baldus Analog nitrous oxide sedation system. This video shows compatible suction connections for both small and large chair suction outlets, adapter usage, and the correct rotary knob adjustment to achieve optimal suction performance during N2O sedation.